Option Enable Utility

Installing or Removing Options

Installing or Removing Options
There are two types of options:
• Hardware option: This option involves some physical change to the analyzer. The option
• Software option: This option involves loading software to initialize your new option.
Click the arrow in the Select Desired Option box. A list of available options, similar to the
list below, will appear.
010 - Time Domain
UNL - Bias Tees w/Atten
014 - Configurable Test Set
016 - Receiver Attenuators
080 - Frequency Offset Mode
081 - Reference Transfer Switch
082 - Active Device Meas Appl
083 - Mixer Measurement Appl
??? - Enter Unlisted Option
Select the option that you with to either install or remove, by clicking on it. Click either
Install or Remove, whichever is appropriate.
If the desired option is not available in the list, select Enter Unlisted Option. A dialog box
appears that will allow you to enter the option number. Enter the option number and
follow the instructions on the display.
B-4
enable utility allows you to load the software needed for the
hardware changes. It is necessary to reinstall hardware options if the
A16 test set motherboard is replaced
The option enable utility is part of this process. It is necessary to
reinstall software options if the A10 frequency reference board is
replaced.
PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
Service Guide E8364-90001