Signal Separation Group Operation; Configurable Test Set - Agilent Technologies N5242A Service Manual

2-port and 4-port pna microwave network analyzers (10 mhz–26.5 ghz)
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

Theory of Operation

Signal Separation Group Operation

Signal Separation Group Operation
The signal separation group divides the source incident signals into a reference path and a
test path. Refer to
Figure 5-4 on page 5-18
• The reference signals are transmitted to the receiver group as the R1, R2, R3, and R4
inputs for 4-port models or the R1 and R2 inputs for 2-port models.
• The test signals are transmitted through—and reflected from—the device under test
(DUT) and then transmitted to the receiver group as the A, B, C, and D inputs for 4-port
models or the A and B inputs for 2-port models.
• Control lines to this group are routed from the A19 test set motherboard.
In this section, the following assemblies are described:
A25–A28 Bridges
A29–A32 Test Port Couplers
• Front Panel
Jumpers—Configurable Test Set
A34–A37 65-dB Source Step Attenuators and A38–A41 Bias Tees (Optional)
A46–A49 Source Bypass Switches and A50 Combiner (Optional)

Configurable Test Set

The configurable test set is included in the standard analyzer and allows you to measure
devices with higher power and higher dynamic range limits than an analyzer without the
configurable test set. On 4-port models, twelve signal paths, routed through front panel
SMA jumpers, comprise the configurable test set. On 2-port models, there are six signal
paths routed through front panels jumpers in the configurable test set.
As shown in
Figure 5-4 on page 5-18
installed between the components listed below. Ports 3 and 4 apply only to 4-port models.
• the A25 port 1 bridge and the A23 mixer brick channel R1
• the A26 port 3 bridge and the A24 mixer brick channel R3
• the A27 port 4 bridge and the A24 mixer brick channel R4
• the A28 port 2 bridge and the A23 mixer brick channel R2
• the A25 port 1 bridge and the A29 test port 1 coupler
• the A26 port 3 bridge and the A30 test port 3 coupler
• the A27 port 4 bridge and the A31 test port 4 coupler
• the A28 port 2 bridge and the A32 test port 2 coupler
• the A29 test port 1 coupler and the A23 mixer brick channel A
• the A30 test port 3 coupler and the A24 mixer brick channel C
• the A31 test port 4 coupler and the A24 mixer brick channel D
• the A32 test port 2 coupler and the A23 mixer brick channel B
5-16
and
Figure 5-5 on page
and
Figure 5-5 on page
PNA Series Microwave Network Analyzers
5-19.
5-19, these jumpers are
Service Guide N5242-90001