Repair and Replacement Procedures

Removing and Replacing the A18 MA 20

Removing and Replacing the A18 MA 20
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-13
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
analyzer bottom side up and raise the receiver deck as shown. Refer to
Removing the Receiver Deck" on page 7-8
3. With a 5/16-inch torque wrench, disconnect the two semirigid cables (item
A18 MA 20. It may be necessary to remove these cables completely to allow the
A18 MA 20 to be removed from the analyzer.
4. Disconnect the ribbon cable (item
5. With a T-10 TORX driver, remove the four mounting screws (item
the analyzer.
Be careful not to damage the center pins of the semirigid cables. Some flexing
CAUTION
of the cables is necessary to remove the assembly. Do not over-bend the
semirigid cables.
6. Remove the A18 MA 20 from the analyzer. Observe the
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-32
for this procedure.
"Removing the Covers" on page
) at the A18 MA 20.
Table 7-2 on page
PNA Series Microwave Network Analyzers
if necessary.
CAUTION
7-70.
Service Guide E8364-90001
E8362A, E8363A, E8364A
7-6. Position the
"Raising and
) from the
) from the side of
above.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents