Repairing and Recovering Option Data

License, option, and model number data are stored in an EPROM on the A10 frequency
reference board and written into the gen.lic file on the hard disk drive. If the data stored
in either of these items is lost, it can be recovered from the other item.
If the data on both the hard disk drive and the frequency reference board is lost, it can not
be recovered. Contact Agilent for assistance. Refer to
Recovery of Data After Repair
• If the A10 frequency reference board is replaced, use the Repair selection to recover
data that has been lost as a result of the repair:
1. Select Repair from the Option Enable menu bar (see
2. Click the Freq Ref board has been replaced check box.
3. Click Begin Repair. The data is written from the gen.lic file into the EPROM.
• If the hard disk drive is replaced, a new gen.lic file is automatically created when the
Network Analyzer application starts, by retrieving the data from the EPROM.
Recovery of Data if Option or Model Numbers are Incorrect
If the analyzer option or model numbers are not listed correctly on the analyzer display, in
either the Option Enable Dialog Box or the About Network Analyzer display, you should
regenerate the options license file, gen.lic:
1. Exit the Network Analyzer application.
2. Remove (or rename) the existing gen.lic file:
a. Open Windows Explorer and navigate to C:\Program Files\Agilent\Network
Analyzer.
b. Delete (or rename) gen.lic.
3. Exit the Windows Explorer application.
4. Open the Network Analyzer application. The application will generate a new gen.lic
file when it starts.
5. Check the option listing:
a. On the System menu, point to Service, and then click Option Enable, or
b. On the Help menu, click About Network Analyzer.
6. If the options are still not listed correctly, contact Agilent for assistance. Refer to
"Contacting Agilent" on page
Service Guide E8364-90001
2-9.
Repairing and Recovering Option Data
"Contacting Agilent" on page
Figure
B-1).
Option Enable Utility
2-9.