Frequency
Tools required
Parts required
CA UT IO N
Figure 16
Repairing the Autosampler

Interface Board

Interface Board
At installation or when defective.
Flat-head screwdriver.
Interface board, see "Optional Interface Boards" on page 228.
The interface board is sensitive to electrostatic discharge. Always use the ESD
strap when handling electronic boards.
1 Switch off the autosampler at the main power switch.
2 Disconnect cables from the interface board connectors.
3 Loosen the screws. Slide out the interface board from the autosampler.
4 Install the interface board. Secure the screws.
5 Reconnect the cables to the board connectors
Exchanging the Interface Board