Exchanging the Needle Assembly
When
When the needle is visibly damaged
When the needle is blocked
Tools required
Description
Wrench, 1/4 inch (supplied in HPLC Tool-Kit)
Hexagonal key, 2.5 mm (supplied in HPLC Tool-Kit)
Pair of pliers
Parts required
#
1
OR
1
Preparations
• Select Start in the maintenance function Change Needle (see
• When the needle is positioned approx.15 mm above the needle seat, remove the front cover.
Personal injury
WA R N I N G
To avoid personal injury, keep fingers away from the needle area during autosampler
operation.
➔
➔
Agilent 1260 Infinity Standard Autosampler User Manual
p/n
G1313-87201
G1313-87202
Do not bend the safety flap away from its position, or attempt to remove the safety
cover.
Do not attempt to insert or remove a vial from the gripper when the gripper is
positioned below the needle.
Description
Needle assembly
Needle assembly (900 µL)
Maintenance
Exchanging the Needle Assembly
"Change Needle"
8
on page 60).
111