Repairing the Sampler

Exchanging Internal Parts

The following procedures require opening the main cover of the sampler. Always
WA RN ING
ensure the sampler is disconnected from the line power when the main cover is
removed. The security lever at the power input socket prevents the autosampler
cover from being taken off when line power is still connected.
The power supply still uses some power, even if the power switch on the front panel
WA RN ING
is turned off. To disconnect the sampler from line power, unplug the power cord.
When opening capillary or tube fittings solvents may leak out. Please observe
WA RN ING
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
The electronics of the sampler will not allow operation when the top cover and the top foam
NO TE
are removed. A safety light switch on the main board will inhibit the operation of the
sampler. Always operate the sampler with the top foam and top covers in place.
Internal components may be sensitive to electrostatic discharge (ESD). Always use an
CAU TI O N
ESD strap when handling internal components (see
page 107).