Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform a reflection one-port calibration. A one port
calibration uses known standards to correct for directivity, source match, and
frequency response errors in narrowband measurements.
To perform a reflection one-port calibration you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
for 50 62 type-N female connector calibrations
HP 850323
for type-N female or type-N male 50 61 connector calibrations
HP 85032B
for 75 61 type-N female connector calibrations
HP 850363
for type-N female or type-N male 75 62 connector calibrations
HP 85036B
for 3.5 mm female or 3.5 mm male 50 0 connector
HP 85033D
calibrations
for Type-F 75 D connector calibrations
HP 85039B
N O T E
If you are going to be using calibration standards other than the default [female type-N), you must
select the type of connector type by pressing (CAL)
Tel:0755-83589391
Measuring Reflection Response
Kit and then selecting the appropriate
Cal
Fax:0755-83539691
Making Measurements
3 - 2 5
Website: www.Sengt.c