深圳市盛腾仪器仪表有限公司
This procedure shows you how to configure the network analyzer for
measurements.
1.
Press (PRESET). Presetting the instruments puts it into a known state with
predefined parameters.
2.
Press @?%I) and then use a softkey to select the type of device that you
will be measuring (amplifier, filter, broadband passive device, mixer, or
cable).
3.
Connect your test device to the network analyzer.
4.
Use the softkeys to select the type of measurement you want to make:
Press Trans3missn if you want to measure the transmission
l
characteristics of an amplifier, filter, or broadband passive device.
Press Reflection if you want to measure the reflection characteristics
l
of your device.
Press Power if you want to measure the RF power of a device. (The
l
Power selection is under the Amplifier menu.)
Press Conwersion Loss if you want to measure the conversion loss of
l
a device. (The Conoersian Loss selection is under the Mixer menu.)
AM
Press
Delay (Option 1DA or 1DB only) if you want to measure the
l
delay of a device. (The AM Delay selection is under the Mixer menu.)
Press SRL (Option 100 only) if you want to measure the structural
l
return loss of a cable. (The SRL selection is under the Cable menu.)
. Press Fault Location (Option 100 only) if you want to measure the
cable fault location. (The Fault Location selection is under the Cable
menu.)