Using Instrument Functions
Using Markers
To Use Multi-Peak or
Multi-peak and multi-notch searches are designed for use when measuring
Multi-Notch Search
multi-pole Elters. Both searches automatically search the measurement trace
from left to right, and position a marker at each local maximum or minimum.
Up to eight maximums or minimums will be found. Searches are limited to
responses above -60 dB. With marker tracking ON, the multi-peak/notch
search will be performed after each sweep.
A maximum (or minimum) point is detected whenever an amplitude
excursion greater than half of a division occurs. The half of a division
excursion requirement must be satisEed on both sides (left and right) of the
peak (or minimum). The maximum (or minimum) point must be >-60 dB.
See Figure 4-8.
4-16
深圳市盛腾仪器仪表有限公司
N O T E
During a multi-peak or multi-notch search, eight markers are always placed on the trace, regardless of
how many peaks or notches are found. The markers that aren't placed on a peak or notch are placed
at the far right of the display on the current maximum frequency point.
Peak Search
Criteria
Figure 4-g. Peak and Minimum Search Criteria
Tel:0755-83589391
x A
/
b
-
X Does Not Meet
Peak Search
Criteria
Fax:0755-83539691
I
-
Division