When fast recall is used in conjunction with a switch connected to the
USER TTL IN/OUT rear panel connector, you can cycle through up to seven
instrument states in sequence by activating the switch.
1. Connect a switch to the USER TTL IN/OUT rear panel connector as shown
in Figure 7-7.
Figure 7-7. Connect a Switch to the USER TTL INIOUT Connector
2. Make sure the analyzer is configured to use the USER TTL
IN/OUT connector for softkey sequencing: press
3. With fast recall toggled to ON, press the switch several times while
observing the analyzer.
4. Notice that with each press of the switch, the hles are highlighted (boxed)
in succession, and that after the last available hle has been used, the
sequence starts again at the top of the softkey menu.
Tel:0755-83589391
Automating Measurements
Measurement Setup and Control with Fast Recall
Fax:0755-83539691
)
OPTIONS
7-43