Dithering to Shift Spurs
C A U T I O N
Activating Spur
Avoidance
C A U T I O N
深圳市盛腾仪器仪表有限公司
Dither shifts all spurs by a small amount once, thus it imposes no sweep time
penalty. But some spurs occurring within the measured frequency band
may not be shifted out of band, and others may be shifted in. Therefore.
dither is most effective for narrowband measurements with a user defined
measurement calibration. lb activate dithering:
2. Make a user-dehned measurement calibration. Refer to Chapter 6
"Calibrating for Increased Measurement Accuracy, ' for calibration
procedures.
The measurement calibration must be performed with the same spur avoid
option used in the measurement or your results may be invalid.
When you activate spur avoidance the analyzer sweeps to a point before a
spur, stops the sweep, shifts the spur, sweeps through the spur location, then
shifts the spur back and continues the sweep. The analyzer determines which
spurs need to be avoided with an algorithm based on frequencies, number of
points, sweep time, and system bandwidth.
1. Press (MENU_) Spur Avuid Opt ions Spw Amid .
2. Make a user-dehned measurement calibration. Refer to Chapter 6
"Calibrating for Increased Measurement Accuracy, ' for calibration
procedures.
penalty can be reduced by setting the start frequency of the measurement as high as possible.
The measurement calibration must be performed with the same spur avoid
mode used in the measurement or your results may be invalid.
Tel:0755-83589391
Fax:0755-83539691
Optimizing Measurements
Reducing Trace Noise
5-15