Display Menus
Run Mode

Run: Log F.E.

3-8
TVA2020 Instruction Manual
FID
FID: 24950ppm
SAMPLING:
Exit=Cancel
Press ENTER or initiate the logging mode from the enhanced probe to
start the VOC logging sequence. After the instrument counts down for the
time selected in the SETUP/Log menu, the word Save appears. When
confirmed, the instrument logs into memory either the highest average or
last reading achieved during the countdown sample. After the information
is stored into memory, the display returns to live measurements and
prompts for a new tag. Each VOC log must be done manually and
individually.
FID
FID: 24950ppm
Save above smp1?
1=Yes
2=Again
The operation of the Log F.E. mode is menu driven from the hand-held
probe display and requires the use of the optional TVA2020 Enhanced
Probe. Whenever the TVA2020 is in the RUN mode with Log F.E. as the
logging selection, the instrument logs its readings based on a pre-
configured route file.
Use the following procedure for use with Fugitive Emissions monitoring.
1. Create a monitoring route using a Fugitive Emissions personal
computer software program as described in Appendix B,
Formats".
2. Copy the route file to the TVA2020 by overwriting the ROUTE.TXT
file in the instrument flash with the one created. It must be named
ROUTE.TXT for the instrument to recognize it. After exiting the USB
mode, the instrument reads the properly named files from flash into the
processor memory.
PID:
FID: 24950ppm
3 SEC
SAMPLING:
Exit=Cancel
PID:
FID:
Save above smp1?
1=Yes
PID/FID
1.1ppm
3 SEC
PID/FID
1.1ppm
2.49%
2=Again
"Text Data
Thermo Fisher Scientific