Cleaning an FID
Cartridge
Thermo Fisher Scientific
Use the following procedure to remove and clean the FID cartridge
frequently
(Figure
5–1). Replace when needed.
1. Remove the hydrogen tank. Turn the instrument off.
2. Using the special spanner wrench provided with the instrument tool
kit, unscrew the black cap holding the FID cartridge in place.
Figure 5–1. Replacing the FID Cartridge
3. Screw the special extractor tool provided with the accessory kit into the
off-center hole in the cartridge
4. Remove the cartridge by pulling on the extractor. Unscrew the
extractor from the cartridge. Clean or replace the cartridge.
5. To insert a new cartridge, reverse the procedure.