Servicing
Replacing the Internal Probe Connector Cable
Replacing the
Internal Probe
Connector Cable
7-28
TVA2020 Instruction Manual
Use the following procedure to replace the internal probe connector cable
(Figure
7–24).
1. Locate the probe connector connection.
2. Disconnect the probe cable from the main board.
3. Loosen the cross recess screw holding the connector assembly in the
housing.
4. Install the new probe connector cable assembly and secure in the
housing.
5. Reconnect the cable to the main board.
Connector on
Main Board
Figure 7–24. Replacing the Internal Probe Connector Cable
Screw