Replacing the
Internal Battery
Charger
Connector Cable
Thermo Fisher Scientific
Use the following procedure to replace the internal battery charger
connector cable
(Figure
1. Locate the battery charger connection.
2. Disconnect the charger cable from the main power cable.
3. Remove the two screws holding the connector in the housing.
4. Re-use the screws to mount the new connector cable.
5. Reconnect the main power cable to the new charger cable.
Screws
Figure 7–23. Replacing the Internal Battery Charger Connector Cable

Replacing the Internal Battery Charger Connector Cable

7–23).
Servicing
Battery Charger Connector from
Main Power Cable
TVA2020 Instruction Manual
7-27