Download Print this page

Thermo Scientific TVA2020 Instruction Manual page 72

Toxic vapor

Advertisement

Display Menus
Memory
3-32
TVA2020 Instruction Manual
a. ROUTE.TXT (Route List)
b. CONFIG.TXT (TVA2020 Configuration)
c. CAL.TXT (TVA2020 Calibration Parameters)
d. FACTORY.TXT (TVA2020 Factory Calibration)
Note TVA2020 only recognizes the above naming format, otherwise the
file will be deleted.
Upon exiting the USB MODE screen, any files that have changed will be
loaded into the instrument memory. If the file format is incorrect, the
following message will appear.
=== WARNING! ===
File read error:
check err file!
Exit=Clr
If you choose Exit to clear, the following displays:
Error in files
1=USB mode to
fix files
2=Revert to old
If you choose to fix the files, re-establish the PC USB connection and open
the error file that was created to identify the line(s) of text in error.
If you choose to revert to the existing files, your changes will be lost unless
they were saved on your PC. The following screen appears:
Reverting
Please wait
Text File Formats
Messages to be uploaded and downloaded must be written in straight
ASCII text and must contain blank spaces between text, not tabs.
For more information, See Appendix B,
"Text Data
Formats".
Thermo Fisher Scientific

Hide quick links:

Advertisement

loading