Calibration - Thermo Scientific TVA2020 Instruction Manual

Toxic vapor

Advertisement

Calibration

Thermo Fisher Scientific
Successful download of the calibration file resets the background
concentration to zero.
The calibration is stored as CAL.TXT in the following format:
Format
CALIBRATION DATA<CR><LF>
VER=xx.xx
<CR><LF>
<CR><LF>
<detector calibration>
<detector calibration>
END<CR><LF>
The detector calibration format is as follows (note: from 1 to 9 span gas
points are allowed):
ddd CALIBRATION:<CR><LF>
<CR><LF>
CALIB
KNOWN
TYPE
CONC
----- ---------- ------------------- -------------------------<CR><LF>
ZERO
aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
SPAN1 aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
.
.
.
. .
.
.
.
. .
.
.
.
. .
SPANn aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
<CR><LF>
<CR><LF>
where:
ddd = detector type
"FID"
"PID"
aaaaaa = gas concentration, floating point, right justified,
leading spaces
uuu = concentration units:
"PPM", parts-per-million
"% ", percent
dd = day of calibration, 01-31
mmm = month of calibration, Jan-Dec
yy = year of calibration, 00-99
hh = hour of calibration, 00-23
mm = minute of calibration, 00-59
ss = second of calibration, 00-59
bbbbbbbb = detector reading for calibration gas, integer,
right justified, leading spaces
fffffffff = count status:
"OK
", no errors
"DET_OFF ", detector reading invalid
Comments
Title, not processed
xx.xx=ver, right just, lead spaces
Blank line, not processed
Blank line, not processed
PID if available, see below
FID if available, see below
Marks end-of-text
CALIB
MEASURED<CR><LF>
DATE/TIME
CALIB VALUE<CR><LF>
.
. . .
.
.
. . .
.
.
. . .
.
Text Data Formats
Calibration
.
.
.
.
.
.
.
.
.
.
.
.
TVA2020 Instruction Manual
B-11