Calibration
Thermo Fisher Scientific
Successful download of the calibration file resets the background
concentration to zero.
The calibration is stored as CAL.TXT in the following format:
Format
CALIBRATION DATA<CR><LF>
VER=xx.xx
<CR><LF>
<CR><LF>
<detector calibration>
<detector calibration>
END<CR><LF>
The detector calibration format is as follows (note: from 1 to 9 span gas
points are allowed):
ddd CALIBRATION:<CR><LF>
<CR><LF>
CALIB
KNOWN
TYPE
CONC
----- ---------- ------------------- -------------------------<CR><LF>
ZERO
aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
SPAN1 aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
.
.
.
. .
.
.
.
. .
.
.
.
. .
SPANn aaaaaa uuu dd mmm yy hh:mm:ss bbbbbbbb COUNTS fffffffff<CR><LF>
<CR><LF>
<CR><LF>
where:
ddd = detector type
"FID"
"PID"
aaaaaa = gas concentration, floating point, right justified,
leading spaces
uuu = concentration units:
"PPM", parts-per-million
"% ", percent
dd = day of calibration, 01-31
mmm = month of calibration, Jan-Dec
yy = year of calibration, 00-99
hh = hour of calibration, 00-23
mm = minute of calibration, 00-59
ss = second of calibration, 00-59
bbbbbbbb = detector reading for calibration gas, integer,
right justified, leading spaces
fffffffff = count status:
"OK
", no errors
"DET_OFF ", detector reading invalid
Comments
Title, not processed
xx.xx=ver, right just, lead spaces
Blank line, not processed
Blank line, not processed
PID if available, see below
FID if available, see below
Marks end-of-text
CALIB
MEASURED<CR><LF>
DATE/TIME
CALIB VALUE<CR><LF>
.
. . .
.
.
. . .
.
.
. . .
.
Text Data Formats
Calibration
.
.
.
.
.
.
.
.
.
.
.
.
TVA2020 Instruction Manual
B-11