Run Mode - Thermo Scientific TVA2020 Instruction Manual

Toxic vapor

Advertisement

Display Menus

Run Mode

Run Mode
3-4
TVA2020 Instruction Manual
How the instrument functions in the RUN mode is governed by the
selection made in the SETUP/Log menu. The LOG menu selects how data
is stored in the instrument memory. There are four possible choices in the
LOG mode as defined in
optional Bluetooth module is active, data automatically streams out
continuously from the instrument. Data will only stream out of the
instrument while in the RUN mode. The data will always stream at in the
same format independent of the log configuration. (See
Communications" in the Options section for more information.)
The TVA2020 performs accurately only after it has been properly set up
(configured).
The RUN mode displays (as governed by the LOG Selection) are:
FID-NONE
FID:
1376ppm
Exit=Stop
FID-AUTO, VOC, or F.E.
FID:
1376ppm
Sampling:
Exit=Stop
You should install the hydrogen tank and wait 2-3 minutes before entering
the RUN mode. Upon entering the RUN mode, the pump will
automatically turn ON and the FID will ignite.
Note For best performance, the instrument must be ON and warmed up
for approximately 30 minutes with the FID flame ignited throughout the
warmup period. If the instrument is PID equipped and the PID is to be
used, the lamp must be ON throughout the warmup period.
"Log
Methods" on page 3-15. In addition, if the
PID/FID-NONE
PID:
FID:
Exit=Stop
PID/FID-AUTO, VOC, or F.E.
PID:
FID:
3sec
Sampling:
Exit=Stop
"Bluetooth
1042ppm
1376ppm
1042ppm
1376ppm
3sec
Thermo Fisher Scientific