11. Place the spacer around the cell assembly
Figure 7–13. Placing the Spacer
12. Place the filter media over the cell assembly centered on top of the
sensor. Make sure the filters are installed in the correct order. Filter
media #2 first, then filter media #1 on top, with the shiny side up
(Figure
7–14).
Figure 7–14. Placing the Filter Media
13. Align the cap key with the notch on the housing. Starting at the side
opposite the notch, press down until the filter cap snaps on to the
housing. If the cap key is incorrectly aligned, there will be a noticeable
bulge on the side of the cap
Figure 7–15. Aligning the Cap Key
(Figure
7–13).
(Figure
7–15).
TVA2020 Instruction Manual
Servicing
Cleaning the PID Lamp