When you press 2=Span, the display will show:
FID
FID:
44750
Span Pt1 (Up/Dn)
Enter=New calib
If more than one reference span gas concentration has been configured,
pressing the Up/Dn keys will allow you to scroll through all defined
reference gas concentration points.
2. To perform the actual SPAN calibration (for example at 100 ppm
isobutylene):
FID-only: Press ENTER
PID/FID: Press 1, 2, or 3
FID
Apply span gas1
100.0ppm
Enter=Start
3. Apply the appropriate span gas to the probe at ambient pressure (using
a clean and labeled gas sampling bag) and then press ENTER.
FID
FID:
44750
Wait for span...
Exit=Cancel
The instrument analyzes the span sample.
When only one reference span gas concentration is used, the ACCEPTED
message appears for a short time and is then replaced by the normal
CALIBRATION menu.
Defining the Response Factor
PID/FID
PID:
4000
FID:
44750
Span Pt1 (Up/Dn)
1=Both 2Pid 3Fid
PID/FID
Apply span gas1
100.0ppm
Enter=Start
PID/FID
PID:
4000
FID:
44750
Wait for span...
Exit=Cancel
TVA2020 Instruction Manual
Calibration
4-17