Cleaning the FID
Detector Cap
Thermo Fisher Scientific
Use the following procedure to clean the FID detector end cap
3).
1. Remove the hydrogen tank from the top of the instrument and turn off
the power.
2. Using the special spanner wrench provided with the tool kit, unscrew
and remove the FID detector cap.
3. Clean the cap using the isopropyl alcohol followed with a de-ionized or
distilled water rinse. Blow out carefully with compressed dry air.
4. Replace the cap.
Figure 5–3. Cleaning the FID Detector Cap