9. The copying window should appear while the file transfers.
10. After transfer, the 'Removable Disk' window will close and the
instrument will automatically reboot to the main menu.
11. If the instrument does not reboot to the main menu, perform the
following:
a. With the USB cable still connected to the computer, disconnect
and reconnect at the instrument. This will turn the instrument off.
b. Press the power key to turn the instrument on.
c. The instrument will display two black lines; however the
bootloader is still active.
d. Return to step 7 in this section and continue from there.
e. If the problem persists after several attempts, contact the Thermo
Fisher Scientific Service Department.
12. Confirm that the calibrations (gas and flow) are still valid. If needed,
they can be transferred back into the instrument using the
Memory/USB Mode functions.
Display Menus
Memory
TVA2020 Instruction Manual
3-35