Run: Log Voc - Thermo Scientific TVA2020 Instruction Manual

Toxic vapor

Advertisement

Display Menus
Run Mode

Run: Log VOC

3-6
TVA2020 Instruction Manual
FID
FID: 24950ppm
A ---------------
0=1
1=A 2=B 3=C
Once the tag information is entered, press the ENTER key. The screen
then changes to the next display. The instrument is now ready to start
logging. The screen displays live readings and the tag name. However, no
logging takes place until you enter the logging mode from the enhanced
probe or press the ENTER key on the instrument.
FID
FID: 24950ppm
BLD01SITE5
Enter=Start log
Press ENTER or initiate the logging mode from the enhanced probe to
start the automatic logging. The instrument counts down from the time
selected in the SETUP/Log menu and stores the reading at the end of the
count down. For instance, if 12 seconds was selected as the logging time,
the instrument will count down from 12 and display
11,10,9,8,7,6,5,4,3,2,1,STORE. The instrument logs into memory either
the highest, average, or last reading during the countdown, as configured in
the SETUP/Log menu.
FID
FID: 24950ppm
Sampling:
Exit=Cancel
This action continues until the instrument shuts down due to low battery,
or the EXIT key is pressed, or the log memory is full.
Whenever the TVA2020 is in the RUN mode with Log VOC as the
logging selection, the instrument operates as a manual survey tool that logs
its readings into memory whenever you initiate logging. When the 1=Run
PID/FID
PID:
FID:
A ---------------
0=1
1=A 2=B 3=C
PID/FID
PID:
FID:
BLD01SITE5
Enter=Start log
PID/FID
PID:
FID:
3sec
Sampling:
Exit=Cancel
1.1ppm
2.49%
1.1ppm
2.49%
1.1ppm
2.49%
3sec
Thermo Fisher Scientific