Display Menus

Memory

Memory
3-30
TVA2020 Instruction Manual
PID span calibration: Date and time of last calibration
FID span calibration: Date and time of last calibration
Note To view calibration information, you must return to the calibration
menu in SETUP. Press EXIT to return to MAIN menu.
This menu allows you to link the TVA2020 to a personal computer (PC)
through the USB port to copy files from and into the instrument, clear data
memory, and update instrument firmware.
Note The Flash memory device is erased prior to the file creation process
to ensure sufficient memory exists to create the data files. Do not store any
files on the instrument Flash drive as they will be lost.
WARNING Charger and USB connections are not allowed in an area
classified as hazardous due to presence of flammable gases or vapors.
To enter the MEMORY menu from the MAIN menu, press
4=MEMORY. The following display will appear:
MEMORY
1=USB Mode
2=Clear Log Mem
3=UpdateFirmware
Choosing 1 = After choosing 1 the display will show "Creating Files Please
wait" as shown below. The file creation can take up to one minute,
depending on the size of the log file.
Creating Files
Please Wait
Do not connect the USB cable to the instrument until the instrument
displays the screen below.