Log - Thermo Scientific Orion 8030cX Instruction Manual

Silica analyzer
Table of Contents

Advertisement

instrument.
Digital Inputs: digital signals can be wired into the analyzer
as inputs, which are used to turn the functions ON/OFF
with preset delay. Functions include Measurement, Deep
Clean, Calibration, Validation and Abort or None. When
they are set to the corresponding function, the
corresponding function is started when there is an input.
The Delay is used to set a specific time delay after input
and then execute
the corresponding function. This
instrument provides two digital inputs, which can be
configured separately.
Analog Outputs: The measurement data can be outputted
to PLC or DCS via 0/4-20mA analog signal . They can be
set to OFF or 4-20mA or 0-20mA. When set to 4-20mA,
the corresponding values of Lower Limit for 4mA and
Higher Limit for 20mA need to be set; when set to 0-20mA,
the corresponding values of Lower Limit for 0mA and
Higher Limit for 20mA need to be set. The lower and higher
limits is set to the lower and upper limits of the
measurement range by default. This instrument provides
two analog outputs, which can be configured separately.
Modbus (RS485): Modbus R485 can be used to output
data via digital signal. RS485 is used to set the station
Address, Baud rate, Data bits, Stop bits and Parity.

Log

Log is used to query various records, including alert/ alarm
log, measurement log, calibration log and operation log.
Click each log to enter different log interfaces.
USB sticks supported: USB devices of fat32 file system. It
is recommended to use standard USB devices of Thermo
Fisher Scientific or USB devices of SanDisk and Kingston
brands.
Alert/Alarm Log
Alert/Alarm logs are used for various alarms and alerts
events occurring during the operation process of the
instrument, including the time, date, event, status of event
and alarm code. When checking alarm records, you can
check based on time (including 1 week, 1 month, 3
months, 6 months, 1 year and all). Alarm records can be
exported by authorized person only. When exporting, a
USB storage device is required to be connected to the
32 | Chapter 5
System Running
8030cX Silica Analyzer