Servicing
Cleaning the PID Lamp
7-10
TVA2020 Instruction Manual
Figure 7–4. Removing the Filter Cap
3. With fine-tipped tweezers, remove both filter media and set aside
(Figure
7–5).
Figure 7–5. Removing the Filter Media
4. Using the X-Acto blade, remove the spacer and set aside
Figure 7–6. Removing the Spacer
5. With fine tipped tweezers, carefully remove the cell assembly by prying
under the cell's edge where connector pins are located
(Figure
7–6).
(Figure
7–7).