Hardware and Startup

Quick Start Procedure

Quick Start
Procedure
2-10
TVA2020 Instruction Manual
CONTROL MENU
1=BT off
2=GPS off
3=More
After running the initial startup diagnostic, the Control menu can be
accessed at any time.
Before starting the unit, perform the following steps:
1. Charge battery.
2. Connect sample probe.
3. Fill/install hydrogen tank (which starts hydrogen flowing).
To start the unit, execute the following procedure:
4. Press ON until the speaker sounds and the display flashes
(approximately 1 second).
5. Press CONTROL.
6. Press 3 to ignite.
7. Press 2 = Setup.
8. Press 1 = Calibrate.
9. Press 6 = Configure.
10. Press 2 = Span Concentration.
11. Enter Span Concentration for calibration gas being used and press
ENTER to accept.