External Influences; Table 1-2. Tva2020 External Influences - Thermo Scientific TVA2020 Instruction Manual

Toxic vapor

Advertisement

External
Influences

External Influences

Ambient temperature
Ambient pressure
Relative humidity
Radiated Susceptibility
Conducted susceptibility
Conducted emission
Radiated emission
ESD Sensitivity
Battery charger supply
voltage
Battery Charger supply
frequency
Thermo Fisher Scientific
Bluetooth
Wireless communications module certified to Bluetooth® ver2.0
Support for SPP (Serial Port Profile)
Transmit Class 2 (typical range of 6')
"This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received above the radiated susceptibility limits,
including interference that may cause undesired operation."
This device contains FCC-ID POOWML-C40
GPS
Frequency
Channels
Update rate
This product is intended for use in indoor and outdoor environments as a
portable instrument carried by a user, as specified in
environmental conditions also apply to the sample stream being monitored.
Table 1–2. TVA2020 External Influences
Reference Operating
Normal Operating
Conditions
Conditions
23 ± 2 °C
-10 to +45 °C
860 to 1060 mbar
70 to 108 kPa
50% ± 10%
FID: 20 to 95%
PID: 20 to 70%
noncondensing
None
Not applicable, battery operated
Not applicable, battery operated
120 or 230 ±1% Vac
100 to 240 Vac
50/60 Hz
50/60 Hz
±0.5 Hz
1575 MHz
up to 12 simultaneously
1 Hz
Operative Limits
-10 and +45 °C
15 and 95%
noncondensing
3 V/m
80 to 2700 MHz
See Part 15 FCC/
Bluetooth requirement
>6000 Volts
100 to 240 Vac
50/60 Hz
Introduction
External Influences
Table
1–2. The same
Transportation and
Storage Limits
-20 to +60 °C
-4 and 140 °F
20 to 108 kPa
0 to 100%
Not applicable
Not applicable
TVA2020 Instruction Manual
1-11