Measurement Of The Peak Factor (Cf); Cf Display Screen (3U) - AEMC 8335 Power Pad User Manual

3-phase power quality analyzer
Table of Contents

Advertisement

Item
Function
Instantaneous value cursor; change values with the cursor using the ◄ and ► buttons
1
2
Harmonic distortion rate for each curve
Instantaneous value of the signals at the intersection of the cursor and the curves
t: time relative to the start of the period (expressed in milliseconds)
3
A1: instantaneous current of phase 1
A2: instantaneous current of phase 2
A3: instantaneous current of phase 3
Note: Filters L1, L2, and L3 display the total current and voltage harmonic distor-
tion for phases 1, 2, and 3, respectively.

5.3.3 Measurement of the Peak Factor (CF)

The CF sub-menu displays the waveforms of the signals measured over one
period and the voltage and current peak factors.

5.3.3.1 CF Display Screen (3U)

This screen displays the phase-to-phase voltage waveforms of one period and the
peak factors.
1
Item
Function
Instantaneous value cursor; change values with the cursor using the ◄ and ► buttons
1
2
Peak factor for each curve
Instantaneous value of the signals at the intersection of the cursor and the graphs
t: time relative to the start of the period (expressed in milliseconds)
3
U1: instantaneous voltage peak factor between phases 1 and 2 (U
U2: instantaneous voltage peak factor between phases 2 and 3 (U
Shop for AEMC products online at:
U3: instantaneous voltage peak factor between phases 3 and 1 (U
Power Quality Analyzer Model 8335
Figure 5-27
PowerMeterStore
www.
2
3
)
12
)
1.877.766.5412
23
.ca
)
31
75