5.1.3 True RMS Current and Voltage

The RMS mode displays the record of the trend of the true half-cycle RMS current
and voltage and the frequency trend curve.

5.1.3.1 RMS Display Screen (3A)

1
2
Item
Function
1
Instantaneous value cursor; change values with the cursor using the ◄ and ► buttons
2
Scale of values in amperes
- Maximum half-period RMS value of the inrush current capture
3
- Indicator of number attributed to the displayed curve
- Value of the measured current
Selection of curves to be displayed:
- 3V: displays the 3 voltages during the inrush current capture
4
- 3A: displays the 3 currents during the inrush current capture
- L1, L2, L3: display the current and voltage in phases 1, 2, and 3, respectively
- Hz: displays the evolution of the network frequency vs time
- t: relative time position of the cursor
5
(t = 0 corresponds to the start of the inrush capture)
- A1, A2, A3: RMS values of currents 1, 2 and 3 at the position of the cursor
6