Menu Options - Promax HD RANGER 2 User Manual

Tv & satellite analyser
Hide thumbs Also See for HD RANGER 2:
Table of Contents

Advertisement

7.8.3

Menu Options

In the bottom of the screen are four menu accessible via the softkeys.
In the Advanced menu there are five parameters to set the attenuation test.
They are:
Threshold Attenuation:
It defines the maximum difference that may exist between the pilot signal of
higher level and the pilot signal of lower level. All pilot signals out of this
range will be removed and not used as a pilot signal during the measurement
process.
Maximum Attenuation:
It sets the attenuation level used by the equipment to show on screen if the
signal passes or fails. When the attenuation level is below this value the
message on screen is "PASS" and when it is above this value is "FAIL".
Pilot 0:
It defines the frequency of the pilot signal 0 (MHz).
Pilot 1:
It defines the frequency of the pilot signal 1 (MHz).
Pilot 2:
It defines the frequency of the pilot signal 2 (MHz).
7-94
Exit: Exits the tool.
Set Reference: Pressing this option the current
power values are captured and they are assigned as
reference values.
Tools: It access the Tools main menu.
Advanced: It access the Advanced menu.