Spo2 System Messages - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

Chapter 10 Pulse CO-Oximetry (SpO2)

SpO2 System Messages

The X Series unit may display the following system messages when monitoring SpO
System Message
INITIALIZING
SEARCHING
CHECK SENSOR
SENSOR FAILURE
CABLE FAILURE
LIFE EXPIRING
LOW PERFUSION
SPO2 DISABLED
TIME LEFT: XX:X HRS
10-16
Cause
The SpO
pulse oximeter is initializing.
2
The unit is searching for a pulse.
The SpO
sensor has become disconnected from the
2
unit, or the sensor is no longer on the patient. Check
sensor and then reconnect it to the unit or reapply it to the
patient.
The SpO
sensor is defective. Replace the sensor.
2
The SpO2 cable is defective. Replace the cable.
Adhesive/Sensor/Cable Failure. The adhesive/sensor/
cable is nearing expiration. Replace the adhesive/
sensor/cable.
Signal too small, move to better perfused site.
A system error has occurred. The X Series unit cannot
take SpO
measurements and should be serviced.
2
For sensors that support expiration, indicates the
approximate time left before sensor expiration.