The Nibp Numeric Display; Nibp Setup And Use - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

Chapter 8 Monitoring Non-Invasive Blood Pressure (NIBP)

The NIBP Numeric Display

When NIBP monitoring has been set up and the X Series unit has begun taking NIBP
measurements, the systolic, diastolic, and mean blood pressure measurements appear on the
NIBP numeric display as follows:
The Artifact symbol (
detects that NIBP measurements (systolic, diastolic, mean) may be inaccurate. The symbol
displays when NIBP measurements are below the specified measurement range for the selected
patient type or when the accuracy of NIBP measurements may be compromised by the presence
of motion artifact, weak pulses, cardiac arrhythmias or other blood pressure artifacts.
Note:
The following sections describe how to set up NIBP monitoring.

NIBP Setup and Use

To take safe and accurate NIBP measurements using the X Series unit, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform NIBP measurements.
1. Select the correct size cuff.
Apply the cuff to the patient.
2.
Connect the inflation hose to the X Series unit and to the cuff.
3.
4.
Configure NIBP alarms and settings (if the current NIBP alarms and settings are not
appropriate).
Press the NIBP button on the X Series unit's front panel and take the blood pressure
5.
measurement.
8-4
) appears on the NIBP numeric display whenever the X Series unit
For a list of system messages that may display on the X Series unit when monitoring
NIBP see "NIBP System Messages" on page 8-15.
www.zoll.com
mmHg
NIBP
138
(112)
89
Artifact Symbol
9650-001355-01 Rev. M

Advertisement

Table of Contents
loading

Table of Contents