Quick Reference Guide Changing Patient Mode 1. Press & to highlight the patient mode field at the top of the display. Press 2. Press & to highlight correct patient mode. Press Neonate: Birth through 28 days, or 44 weeks gestational Pediatric: 29 days through 12 years Adult: 12 years or greater Changing Screen Views...
Quick Reference Guide ECG Monitoring Apply ECG electrodes. 3-Lead Electrode Placement 5-Lead Electrode Placement Changing the Lead Source in Waveform Trace 1 I, II, Press to cycle through leads until the desired lead is displayed. III... Changing the Lead Source in Waveform Trace 1-4 1.
Page 4
Quick Reference Guide 12-Lead ECG Monitoring Apply ECG electrodes. Jugular Notch Angle of Louis Limb Lead Electrode Placement V-Lead Electrode Placement Acquiring a 12-Lead ECG Snapshot 1. Press to enter 12-lead mode. 2. Check for a good quality signal and ensure all leads are displayed.
Page 5
3. Attach the hose to the Propaq MD unit. 4. Apply the cuff to the patient’s limb. 5. Place the limb at the same level as the patient’s heart. 6. Verify the correct patient mode on the Propaq MD unit. Single Lumen Hose Connection Dual Lumen Hose Connection Initiating an NIBP Reading 1.
Page 6
Press 2. Press to highlight Remove. Press Note: You can add the pleth waveform back to the display by selecting Insert from the Source Label menu or by reconnecting the sensor to the Propaq MD unit and the patient.
Page 7
Connecting the CO Sampling Line 1. Slide open the CO inlet port cover on the Propaq MD unit. 2. Insert the fitting at end of the sampling line into the inlet port. 3. Rotate the fitting clockwise to tighten. 4. Attach sampling line to patient or endotracheal tube.
Page 8
Quick Reference Guide Enabling Individual Alarms & Setting Alarm Limits 1. Press & to highlight the parameter label for the alarm you want to set (e.g., HR/PR). Press 2. Press & to highlight the desired parameter alarm (e.g., HR/PR Alarm). Press 3.
Sync mode. Verify sync marker is placed over the R-wave. 2. Press to select desired energy. 3. Press to charge the Propaq MD unit. 4. After the Propaq MD unit is charged, stand clear then press and hold to deliver energy.
Page 10
Quick Reference Guide Starting the Pacer & Changing Pacer Settings 1. Press to bring up the Pacer menu. Pacer 2. Confirm Pacer Mode and Pacer Rate. 3. Press & to highlight Start Pacer. 4. Press to begin pacing. 5. Press &...