Chapter 9 Monitoring CO2
Setting CO
and Respiration Rate Alarms
2
The X Series unit sounds alarms whenever measurements are outside set limits for the
following:
High and Low EtCO
High and Low Respiration Rate (in Breaths/Minute)
High FiCO
No Breath Alarm
Note:
Enabling/Disabling Alarms and Setting CO
You can enable (or disable) CO
Alarms quick access key (
To configure CO
1. Press the More quick access key (
Press
2.
3.
Press the Limits quick access key (
Use the navigation buttons to select appropriate alarm menu. For CO
4.
EtCO2 Alarm, FiCO2 Alarm, or RR/BR Alarm.
On the alarm settings menu, use the navigation keys to select the fields that you want to
5.
change. The fields are
Upper Limit Enable
Lower Limit Enable (only EtCO2 and RR/BR Alarms)
Upper Limit
Lower Limit (only EtCO2 and RR/BR Alarms)
When you are finished changing values on the alarm menu, select the Backarrow key to exit
6.
the menu.
9-8
2
2
Use the Resp Parameter Control Panel to set the duration of the No Breath Alarm, or to
disable it (see "Using the Resp Parameter Control Panel" on page 7-4).
2
) or the CO2 Parameter Control Panel.
alarms through the Alarms quick access key:
2
.
www.zoll.com
Alarm Limits
2
alarms and set upper and lower alarm limits through the
).
).
Limits
, the alarm menus are:
2
9650-001355-01 Rev. M