Chapter 9 Monitoring CO2
Setting CO
and Respiration Rate Alarms
2
The X Series unit sounds alarms whenever measurements are outside set limits for the
following:
High and Low EtCO
•
High and Low Respiration Rate (in Breaths/Minute)
•
High FiCO
•
No Breath Alarm
•
Note:
Enabling/Disabling Alarms and Setting CO
You can enable (or disable) CO
Alarms quick access key (
To configure CO
1. Press the More quick access key (
Press
2.
3.
Press the Limits quick access key (
Use the navigation buttons to select appropriate alarm menu. For CO
4.
EtCO2 Alarm, FiCO2 Alarm, or RR/BR Alarm.
On the alarm settings menu, use the navigation keys to select the fields that you want to
5.
change. The fields are
Upper Limit Enable
•
Lower Limit Enable (only EtCO2 and RR/BR Alarms)
•
Upper Limit
•
Lower Limit (only EtCO2 and RR/BR Alarms)
•
When you are finished changing values on the alarm menu, select the Backarrow key to exit
6.
the menu.
9-8
2
2
Use the Resp Parameter Control Panel to set the duration of the No Breath Alarm, or to
disable it (see "Using the Resp Parameter Control Panel" on page 7-4).
2
) or the CO2 Parameter Control Panel.
alarms through the Alarms quick access key:
2
.
www.zoll.com
Alarm Limits
2
alarms and set upper and lower alarm limits through the
).
).
Limits
, the alarm menus are:
2
9650-001355-01 Rev. M