Constant Metronome Disabled; Fully Release Prompt - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

Chapter 19 Real CPR Help

Constant Metronome Disabled

AED Mode
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval.
If set to Never, the metronome is silent.
Analysis/CPR Protocol or Manual Mode
If Enable Metronome is set to Always, the metronome is active only when the reported
compression rate falls below 80 cpm during the CPR period. If the chest compression sensor
is connected, the metronome is suspended after compressions have not been detected for at
least 2 seconds.
If set to Never or AED Mode Only, the metronome is silent.
The following table indicates when the CPR metronome is active.
Enable
Metronome
Setting
Constant Metronome Enabled
AED Mode
Only
Always
Never
Constant Metronome Disabled
AED Mode
Only
Always
Never

FULLY RELEASE Prompt

The X Series unit can be configured to display the text prompt FULLY RELEASE, which
instructs rescuers to lift (fully release) their hands from the patient's chest after compressions to
allow full recoil. This prompt occurs 45 seconds into the CPR interval.
The
FULLY RELEASE
19-4
AED Mode
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
Off
On at start of CPR period
On at start of CPR period
Off
text prompt is only available when using Adult CPR electrodes.
www.zoll.com
Analysis/CPR
Protocol Mode
Off
On when CPR detected
Off when CPR Idle
Off
Off
On only at <80
compressions per minute
Off
Manual Mode
Off
On when CPR detected
Off when CPR Idle
Off
Off
On only at <80
compressions per minute
Off
9650-001355-01 Rev. M