Configuring NIBP Alarms and Settings

The last step in preparing to perform NIBP measurements is to ensure that the necessary alarms
are enabled (or disabled), that alarm limits are appropriate, and that the NIBP settings are
correct.

Enabling/Disabling NIBP Alarms and Setting Alarm Limits

When enabled, the X Series unit sounds alarms whenever measurements are outside set limits
for the following:
High and Low Systolic Pressure
High and Low Diastolic Pressure
High and Low Mean Arterial Pressure (MAP)
To enable (or disable) NIBP alarms and set upper and lower alarm limits, you can do so through
the Alarms quick access key (
To configure NIBP alarm limits through the Alarms quick access key:
1. Press the More quick access key (
Press the Alarms quick access key (
2.
Press the Limits quick access key (
3.
Use the navigation keys to highlight and select the appropriate alarm menu selection. For
4.
NIBP, the alarm menu selections are NIBP Systolic Alarm, NIBP Diastolic Alarm, and
NIBP MAP Alarm.
5.
On the selected NIBP alarm settings menu, use the navigation keys to select the fields that
you want to change. The fields are
Upper Limit Enable
Lower Limit Enable
Upper Limit
Lower Limit
6.
When you are finished changing values on the alarm settings menu, select the back arrow to
exit the menu.
Setting Upper and Lower Systolic Alarm Limits
Initially, the NIBP Systolic Alarm Settings menu specifies that the NIBP systolic pressure
alarms are enabled (
limits. The upper and lower limits can be
the default NIBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Patient Type
Adult
Pediatric
Neonate
9650-001355-01 Rev. M
) or the NIBP Parameter Control Panel.
).
).
).
Limits
or disabled (
and displays the default upper and lower systolic
)
),
ON
OFF
ON
NIBP Systolic Limit Default
Lower: 75 mmHg
Upper: 220 mmHg
Lower: 75 mmHg
Upper: 145 mmHg
Lower: 50 mmHg
Upper: 100 mmHg
X Series Operator's Guide
Configuring NIBP Alarms and Settings
or
(default is
). The following table lists
OFF
OFF
NIBP Systolic Limit Range
Lower: 30-258 mmHg
Upper: 32-260 mmHg
Lower: 30-158 mmHg
Upper: 32-160 mmHg
Lower: 20-118 mmHg
Upper: 22-120 mmHg
8-11