Chapter 6
Monitoring ECG
This chapter describes how to use the X Series unit to monitor ECG.
X Series units can perform ECG monitoring through 3-, 5-, or 12-lead ECG patient cables,
Multi-Function Pads, or standard defibrillation paddles. The use of an ECG patient cable and
electrodes is required, however, to monitor ECG during pacing.
OneStep cables and electrodes cannot be used for ECG monitoring.
You can use a 3-Lead, 5-Lead, or 12-Lead wire configuration for ECG monitoring (see 12-Lead
ECG Interpretive Analysis for more information on 12-lead monitoring).
9650-001355-01 Rev. M
X Series Operator's Guide
6-1