Chapter 14 12-Lead ECG Interpretive Analysis

12-Lead Interpretive Analysis

After observing the patient's ECG and determining that all 12-lead traces display correctly, you
can initiate 12-Lead Interpretive Analysis (12-Lead Interpretive Analysis must be enabled
through the Supervisor Menu).
Note:
To begin 12-Lead Interpretive Analysis, press the Acquire quick access key (
Patient Info panel displays provide the patient's information by using the navigation keys to
highlight and select a parameter on the Patient Info panel, then press the Select key (see
"Entering Patient Information" on page 14-3). The X Series unit displays the Acquiring 12-
Lead Status bar as it collects 10 seconds of 12-Lead ECG data:
After acquiring the ECG data, the unit saves the data and displays the Saving 12-Lead Status
Bar in the following manner:
14-8
12-Lead Interpretive Analysis is for Adult patients only. It is important that you
provide the patient's age and gender prior to performing ECG analysis using the
Inovise 12L Interpretive Algorithm. Entering the correct age and gender ensures that
the highest degree of ECG analysis is obtained. If you do not enter the patient's age or
gender the X Series uses the default age (45 years) and default gender (male). See
"Entering Patient Information" on page 14-3.