Selecting The Temperature Label - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

Selecting the Temperature Label

Through the Temp Parameter Control Panel, you can configure alarms (as described
previously) and select descriptive Temperature channel labels that will appear on the numeric
display and on the trend report:
Figure 12-2 Temp Parameter Control Panel
You can select one of the following labels for each temperature channel:
Source Label
ART
CORE
ESOPH
RECT
SKIN
VEN
NASO
If you don't select a label, the Temperature channels appear with the default labels of T1 and
T2.
9650-001355-01 Rev. M
Temp
T1 Alarm
T2 Alarm
T1 Source Label
T2 Source Label
Description
Arterial temperature probe
Central body temperature or tympanic membrane temperature
probe
Esophageal temperature probe
Rectal temperature probe
Skin temperature probe (surface application)
Ventilator airway temperature probe
Nasopharynx or nasal/oral temperature probe
X Series Operator's Guide
Selecting the Temperature Label
Lower
Upper
100
90
85
98
5
8
T1
T2
12-5