Using the Heart Rate Parameter Control Panel

To configure alarms through the Heart Rate Parameter Control Panel, use the navigation keys to
highlight and select the Heart Rate meter and display the HR/PR Parameter Control Panel:
Figure 7-2 HR/PR Parameter Control Panel
Selecting the
can enable/disable Heart Rate alarms and set alarm limits.

RESP System Message

When monitoring Respiration using impedance pneumography, the X Series unit may display
the following message:
System Message
RESPIRATION FAULT
9650-001355-01 Rev. M
prompt displays the HR/PR Alarm Settings menu, on which you
HR/PR Alarm
Cause
The impedance measured through the ECG lead is
too high for the unit to detect Respiration changes,
but not high enough to indicate an ECG lead fault
(the impedance may vary between 1800 and 2800
ohms).
X Series Operator's Guide
RESP System Message
7-9