Specifications
Blood Pressure Validation:
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.
NIBP Accuracy:
NIBP accuracy has been verified utilizing clinical test methods which have been determined to
meet the requirements of EN ISO 81060-2:2012.

Invasive Pressures

Number of Channels: 3
Pressure range: -30 to 300 mmHg
Pressure Accuracy: +/- 2 mmHg or 2% of reading, whichever is greater, plus error.
Pulse Rate Range: 25 to 250 BPM
Pulse Rate Accuracy: +/- 3 BPM, or +/- 3% of value whichever is greater
Pulse Rate Display: Average of last 4 beat-to-beat intervals.
Zero Adjust: + / - 200 mmHg
:
Sensitivity: 5uV/V/mmHg
Offset: +/- 125 mmHg including offset
Excitation Impedance Range: 150 to 10,000 ohms
Excitation Voltage: 4.75 +/- 0.25 VDC
Connector: 6-pin circular MS3100 series
Connect to:
Signal Type
A-24
A
B
C
Sig (-)
Exc (+)
Sig (+)
www.zoll.com
D
E
Exc (-)
shield