Pediatric Pacing

Noninvasive pacing of pediatric patients is performed in an identical manner to adult pacing.
Smaller size pediatric therapy electrodes are available for patients weighing less than 33 lbs/15
kg. If it is necessary to pace for more than 30 minutes, periodic inspection of the underlying
skin is strongly advised. Carefully follow all instructions on electrode packaging.
Pace Fault
The X Series unit may display the following messages when pacing.
System Message
PAUSED
PACING
PACING: CHECK THERAPY
ELECTRODES
ECG FAULT: PACING IN FIXED MODE
PACING: SHORT DETECTED
9650-001355-01 Rev. M
Description
The pacer has paused pacing the patient.
The pacer is pacing the patient.
Therapy pads are not connected or are not
making good contact with the patient. Apply
therapy pads to the patient.
The pacer is pacing in Fixed mode because
the ECG lead displayed in the primary
waveform is faulted.
The pacer output is short circuited due a test
plug connection or a device/MFC fault.
X Series Operator's Guide
Pacing in Fixed Mode