Alarms

Heart Rate Alarms:
Audible: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, and a repetition
interval of 15 seconds.
Visual: Heart Rate Alarm causes the heart rate to be displayed in red, with a white background.
The red device status LED will flash a rate of 1.7 Hz.
Lead Fault Alarm:
Audible: 3 pulse, 500 Hz, triplet tone with a PW of 200 msec, a PRI of 200 msec. The lead
fault tone repeats at a repetition interval of 30 seconds.
Visual: Lead Fault condition causes a LEAD FAULT message to be displayed on the trace
along with a dashed line the width of the trace.
Physiological Alarms (NIBP, SpO
Audible: Same as Heart Rate Alarm
Visual: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.
Audio Pause (Silence) Duration: 90 seconds.
Invalid Operation Alert Tone:
A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.
Maximum Alarm Delay (Includes Alarm Condition Delay and Signal
Generation Delay):
Heart rate/pulse rate:
i f source is ECG, 9 seconds
if source is SpO
if source is IBP, 6 seconds
if source is NIBP, no hold off
SpO
, SpCO, and SpMet Saturation: 10 seconds
2
EtCO
: 7 seconds
2
FiCO
: 5 seconds
2
IBP (Systolic, Diastolic, Mean): 3 seconds
Temperature: 2 seconds
A-16
, Resp, CO
2
, 10 seconds
2