Selecting Sphb Venous Mode; Specifying The Sphb Averaging Time; Selecting The Heart Rate/ Pulse Rate (Hr/Pr) Tone; Spo2 System Messages - ZOLL Propaq M Operator's Manual

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Selecting SpHb Venous Mode

If the SpHb option is installed, the Propaq M unit allows you to specify whether to use venous
mode (On) as the blood sample source or not (Off). When Off (default) the Propaq M uses
arterial as the blood sample source.

Specifying the SpHb Averaging Time

If the SpHb option is installed, the Propaq M unit allows you to choose the time period over
which SpHb values will be calculated: Short, Medium, and Long (default). Averaging time
represents the approximate time period the Propaq M will use to calculate SpHb values. Longer
averaging times allow you to see subtle changes in the SpHb value and tend to improve
accuracy.

Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone

The unit allows you enable or disable the tone that the monitor uses to indicate detection of the
patient's pulse:

SpO2 System Messages

The Propaq M unit may display the following system messages when monitoring SpO
System Message
INITIALIZING
SEARCHING
CHECK SENSOR
SENSOR FAILURE
CABLE FAILURE
LIFE EXPIRING
LOW PERFUSION
9650-000820-01 Rev. K
or
(no tone sounds). The default tone is
On
Off
Cause
The SpO
The unit is searching for a pulse.
The SpO
unit, or the sensor is no longer on the patient. Check
sensor and then reconnect it to the unit or reapply it to the
patient.
The SpO
The SpO2 cable is defective. Replace the cable.
Adhesive/Sensor/Cable Failure. The adhesive/sensor/
cable is nearing expiration. Replace the adhesive/
sensor/cable.
Signal too small, move to better perfused site.
Propaq M Operator's Guide
SpO2 System Messages
.
Off
pulse oximeter is initializing.
2
sensor has become disconnected from the
2
sensor is defective. Replace the sensor.
2
:
2
10-15