AED Pro Error Log Messages

The AED Pro unit records internal errors in an error log, which you can retrieve by using a
Windows-based personal computer with ZOLL Administration Software (ZAS).
To retrieve the error log, follow these steps:
Step
1
2
3
4
5
After saving the error log file on the computer, you can use ZAS to clear the error log on the
AED Pro unit.
Table 2-4 lists critical error log messages.
Important: If any of these messages appears in the error log, contact ZOLL Technical Service
Table 2-4. AED Pro Error Log Messages
Error
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
2–6
Action
Turn off the AED Pro unit.
Start the ZOLL Administration Software (ZAS) on the computer.
Position the AED Pro unit so that its IrDA port has a clear line-of-sight with the computer's
IrDA port and the ports are about 10 to 18 inches (25 to 45 cm) apart.
Press and hold the unit's On/Off button for at least 5 seconds to start the unit in nonrescue
mode.
After the unit issues the message COMMUNICATIONS ESTABLISHED, refer to the ZAS
help for instructions on saving the unit's history file and viewing the included error log.
as described on page vii.
Error Message
The patient relay is not working properly
Attempting to charge the capacitor while the SAFELINE in the CPLD is set - should not
charge
The capacitor takes too long to charge
The H-bridge test failed while the bridge was disabled. There should have been 0 current
measured.
VCAP1 and VCAP2 are not within a defined range of each other
The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the positive test on the H-bridge
The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the negative test on the H-bridge
Defib did not charge during the power-on charge test
The Defib task is not receiving A2D vCap1 event samples - needed for charging
FnSafety has determined that the charge has dropped below a specified value during the
holding of charge.
The capacitor failed to charge
The capacitor over-charged
The capacitor contained an unsafe voltage in an idle state