2 Apply ECG Electrodes/Hands-Free Therapy Electrodes

Apply ECG electrodes, attach lead wires, and connect the ECG cable to the X Series side panel
(see Chapter 6, "Monitoring ECG" for instructions on attaching ECG electrodes to the patient).
Attach hands-free therapy electrodes according to instructions on the electrode packaging.
Connect these therapy electrodes to the multifunction cable (MFC).
Note:
Therapy Electrode Application
WARNING!
Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.
1. Apply one edge of the pad securely to the patient.
Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
2.
pockets between the gel and skin.
Ensure that hands-free therapy electrodes are making good contact with the patient's skin and
3.
are not covering any part of any other ECG electrodes.
Press the Lead quick access key and select I, II, or III to provide the largest amplitude QRS
4.
complex.
Note:
5.
Verify that R-waves are being properly detected by confirming that a QRS tone occurs with
each displayed R wave or by verifying that the X Series unit's heart rate display accurately
reflects the patient's pulse rate.

3 Press PACER button

Press the PACER button on the front panel of the unit. The Pacer Settings window displays.
Mode
Rate
Output
9650-001355-01 Rev. M
Do not use ZOLL OneStep cables for pacing.
Pad
Skin
When the Pacer is on, the lead selection is restricted to Leads I, II, or III.
Pacer Settings
Demand
60
70
Start Pacer
Turn Pacer Off
X Series Operator's Guide
Pacing in Demand Mode
18-3