Chapter 21 Patient Data

Storing Data

The X Series unit continuously stores information in a full disclosure case log for the patient
being monitored. The X Series can maintain up to 150 full disclosure cases that contain
treatment events, trends, ECG and other continuous waveforms, monitoring and event
snapshots, and 12-lead snapshots and analysis. The X Series unit can, at a minimum,
concurrently store the following information:
32 monitor snapshots
500 non-ECG events
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance
The actual information that is stored depends on usage. Also, the specific combination of stored
continuous waveform data depends on how the waveform recording settings are configured in
the Supervisor menu.
Note:

Log Capacity Indicator

The Log quick access key has an indicator bar which shows an approximation of the current
level of data storage in the log.
Indicator bar
When data storage reaches its capacity, the unit performs automatic log management by
deleting the oldest full disclosure case. If the log is not cleared or transferred, the X Series unit
continues to delete cases as needed to acquire storage capacity. For more information on
clearing or transferring the log, see page 21-4.

Capturing a Data Snapshot

Press the Snapshot button (
waveform patient data. The unit captures 12 seconds proceeding and 12 seconds following the
button press.
The X Series unit can store a minimum of 32 snapshots, including:
Monitor snapshots
Defibrillator snapshots
Treatment snapshots
Pacer snapshots
Alarm snapshots
Presenting Rhythm snapshots
Note:
21-2
The X Series unit retains stored cases even if you turn off or remove the auxiliary
power adapter.
LOG
The log is full.
LOG
The log is empty.
) on the front panel to capture a 24-second period of numeric and
Presenting Rhythm snapshots are only taken at the start of a new patient. The snapshot
will not be retaken if the unit is shut off for less than two minutes.