Pacer; Co2 - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

DC Auxiliary Operating Power:
Input: 12-24V
Output: 14.45V
IP Rating: IP24

Pacer

Type: External transcutaneous pacing
Pacer Rate: 30 to 180 BPM ± 1.5%.
Output Current: 0 to 140 mA ± 5% or 5 mA (whichever is greater)
Modes: Demand and Fixed
Status Indicators:
ECG lead fault, pace marker on monitor and chart, start/stop indicator on
display.
Pulse Type: Rectilinear, constant current
Pulse Width: 40 ms +/-2 ms
CO
2
Range: 0 to 150 mmHg
Accuracy CO
CO
Partial Pressure*
2
0-38 mmHg
39-99 mmHg
100-150 mmHg
* At sea level.
** Accuracy applies for breath rates of up to 80 bpm. For breath rates above 80 bpm,
accuracy is 4 mmHg or ±12% of reading whichever is greater, for EtCO
18 mmHg. This is tested according to and is compliant with ISO 21647. To achieve the
specified accuracies for breath rates above 60 breaths/minute, the Microstream FilterLine
H Set for Infant/Neonatal must be used. Above 40 C, ± 1mmHg or ± 2.5% (whichever is
greater) has to be added to the tolerance of the accuracy specs.
CO2 Sampling Rate: 50 msec
Drift of Measurement Accuracy: Over any 24-hour period, the accuracy claims listed above
are maintained.
The accuracy specification is maintained to within 4% for the following gas mixtures (all
values are in Vol.%).
CO
2
1.0 to 13 0 to 97.5
9650-001355-01 Rev. M
8300-000006
11A
4.15A
80W peak
:
2
Accuracy **
± 2 mmHg
± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)
± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)
N
O
2
2
0 to 100
X Series Operator's Guide
N
O
H
0
2
2
0 to 80
Dry to saturated
Pacer
values exceeding
2
Anesthetic Agents
According to EN 21647
A-19